Logo Medical Science Monitor

Call: +1.631.470.9640
Mon - Fri 10:00 am - 02:00 pm EST

Contact Us

Logo Medical Science Monitor Logo Medical Science Monitor Logo Medical Science Monitor

01 March 2004

Reliability of the rapid bedside whole-blood quantitative cardiac troponin T assay in the diagnosis of myocardial injury in patients with acute coronary syndrome

Salam Saadeddin, Mohammed Habbab, Hisham Siddieg, Mahmoud Fayomi, Rofaida Dafterdar

Med Sci Monit 2004; 10(3): MT43-46 :: ID: 11602

Abstract

Background: A rapid bedside whole-blood quantitative cTnT assay has recently been developed. We evaluated
the reliability of this test for the diagnosis of myocardial injury in patients with acute coronary syndrome (ACS).
Material/Methods: Whole-blood cTnT levels were measured in 96 patients with ACS using the Roche Cardiac Reader[sup]®[/sup] rapid bedside assay device, and the results were compared with serum cTnT levels in the same patients measured by the Roche Elecsys[sup]®[/sup] Immunoanalyzer. There were 50 patients with clinical evidence of myocardial injury and 56 without.
Results: From the qualitative point of view (reporting negative or positive tests), the results of the rapid bedside tests were identical to those obtained by the serum immunoanalyzer. From
quantitative the point of view, the rapid bedside tests could not measure exact values below
0.1 ng/ml (reported negative) or above 2.0 ng/ml (reported >2.0). The measurements made by the rapid bedside tests within the range of 0.1 to 2.0 ng/ml correlated well with those of the serum immunoanalyzer (Cardiac Reader[sup]®[/sup] cTnT=0.61, Elecsys[sup]®[/sup] cTnT+0.12; r=0.88), but their mean values were significantly lower (1.20±0.71 vs. 1.41±1.03, p=0.0007).
Conclusions: The rapid bedside cTnT assay correlates well with immunoanalyzer measurements between
the values of 0.1 and 2.0 ng/ml. However, they tend to give significantly lower values and fail
to give exact values below 0.1 and above 2.0 ng/ml, which may affect their performance in monitoring and managing patients with ACS, and limit their use in predicting outcome.

Keywords: Coronary Arteriosclerosis - diagnosis, Dose-Response Relationship, Drug, Heart Diseases - diagnosis, Heart Injuries - diagnosis, Immunoassay - instrumentation, Immunoassay - methods, Myocardium - metabolism, Reagent Kits, Diagnostic, Reproducibility of Results, Sensitivity and Specificity, Troponin I - blood, Troponin T - blood, Coronary Artery Disease - diagnosis, Dose-Response Relationship, Drug, Heart Diseases - diagnosis, Heart Injuries - diagnosis, Immunoassay - methods, Myocardium - metabolism, Reagent Kits, Diagnostic, Reproducibility of Results, Sensitivity and Specificity, Troponin I - blood, Troponin T - blood

Add Comment 0 Comments

Editorial

01 April 2024 : Editorial  

Editorial: Forty Years of Waiting for Prevention and Cure of HIV Infection – Ongoing Challenges and Hopes for Vaccine Development and Overcoming Antiretroviral Drug Resistance

Dinah V. Parums

DOI: 10.12659/MSM.944600

Med Sci Monit 2024; 30:e944600

0:00

In Press

05 Mar 2024 : Clinical Research  

Role of Critical Shoulder Angle in Degenerative Type Rotator Cuff Tears: A Turkish Cohort Study

Med Sci Monit In Press; DOI: 10.12659/MSM.943703  

06 Mar 2024 : Clinical Research  

Comparison of Outcomes between Single-Level and Double-Level Corpectomy in Thoracolumbar Reconstruction: A ...

Med Sci Monit In Press; DOI: 10.12659/MSM.943797  

21 Mar 2024 : Meta-Analysis  

Economic Evaluation of COVID-19 Screening Tests and Surveillance Strategies in Low-Income, Middle-Income, a...

Med Sci Monit In Press; DOI: 10.12659/MSM.943863  

10 Apr 2024 : Clinical Research  

Predicting Acute Cardiovascular Complications in COVID-19: Insights from a Specialized Cardiac Referral Dep...

Med Sci Monit In Press; DOI: 10.12659/MSM.942612  

Most Viewed Current Articles

17 Jan 2024 : Review article  

Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron Variant

DOI :10.12659/MSM.942799

Med Sci Monit 2024; 30:e942799

0:00

14 Dec 2022 : Clinical Research  

Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase Levels

DOI :10.12659/MSM.937990

Med Sci Monit 2022; 28:e937990

0:00

16 May 2023 : Clinical Research  

Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...

DOI :10.12659/MSM.940387

Med Sci Monit 2023; 29:e940387

0:00

01 Jan 2022 : Editorial  

Editorial: Current Status of Oral Antiviral Drug Treatments for SARS-CoV-2 Infection in Non-Hospitalized Pa...

DOI :10.12659/MSM.935952

Med Sci Monit 2022; 28:e935952

0:00

Your Privacy

We use cookies to ensure the functionality of our website, to personalize content and advertising, to provide social media features, and to analyze our traffic. If you allow us to do so, we also inform our social media, advertising and analysis partners about your use of our website, You can decise for yourself which categories you you want to deny or allow. Please note that based on your settings not all functionalities of the site are available. View our privacy policy.

Medical Science Monitor eISSN: 1643-3750
Medical Science Monitor eISSN: 1643-3750